Mettler-Toledo Thornton

June 16, 2008
New Suitability Standards Set Tests TOC

System suitability testing is a critical requirement in qualifying instruments used for Total Organic Carbon (TOC) measurements in pharmaceutical applications. The System Suitability Standards Set from Mettler-Toledo Thornton providedes the reference standard solutions necessary to perform testing for the 5000TOC Sensor in conformance with USP  643  and EP 2.2.44 for System Suitability Testing. The System Suitability Standards Set is packaged in 500 mL containers and includes one bottle of 500 ppb p-Benzoquinone, one bottle of 500 ppb Sucrose and two bottles of Reagent Water. The Calibration Standards Set is also packaged in 500mL containers and includes one bottle of 500ppb Calibration Solution as Sucrose and one bottle of 250ppb Verification Solution as Sucrose. Mettler-Toledo Thornton uses only USP Reference Standards, thereby assuring consistent quality and compliance. Reference certificates are provided for documentation support.

Note: System Suitability Standards and Calibration Standards are perishable products. The USP and EP recommend that System Suitability Standards and Calibration Standards must be in a fresh condition; therefore, testing should be performed as soon as possible after the solutions are manufactured.

Related Content