Analytical Advances for Pharmaceutical Manufacturing: More information in Less Time

Nov. 1, 2010
Pharmaceutical manufacturers have to balance several conflicting requirements in their testing and quality assurance programs.

The cost and speed of analysis as well as the amount of information obtained must be balanced to be as cost and time efficient as possible. In addition, one has to work within the constraints imposed by regulating agencies. This often defines what testing is required and what information must be obtained.

Recent advances in analytical instruments have led to the development of techniques that allow us to increase the speed and amount of information collected on pharmaceutical samples. This has come from a couple of recent developments: first, the development of instruments able to do quantitatively measure the heat flow of samples at high scanning rates in heating and cooling. The development of high speed differential scanning calorimetric (DSC) techniques allows the detection of weak transitions to much lower levels than previously possible, allowing the determination of amorphous content to very low levels. In addition, these techniques allow the use of smaller sample sizes and much higher through-put.

Next, the development and commercialization of spectral DSC techniques, where DSC is coupled with either Raman or NIR, allows more precise determination of the material’s state than either technique alone. The use of these techniques, particularly Raman Spectroscopy, allows one to see how the material response to processing conditions simulated in the DSC as well as detect structural changes caused by changes in the process.While evolved gas analysis techniques (EVA) have been available and used for some time now, they have tended to be limited in scope and often custom made systems. Recently we have seen the options in these techniques expand to include infrared spectroscopy (FTIR), mass spectroscopy (MS), and gas chromatography mass spectrometry (GCMS). In addition, EVA techniques have become turn-key and allow the laboratory easily to detect and define residual solvents and contaminants in final products. In the following sections, we will look at these techniques in greater detail.

Fast Scanning Rate DSC: More Information Available Faster.

Originally developed to study polymer crystallization, Fast Scanning Rate DSC has become a powerful tool in pharmaceutical studies. One of its advantages is scanning at rates from 200 to 750 degrees C per minute dramatically shortens the run time for pharmaceuticals and allow for an immense increase in the number of samples run during a shift. At 200 per minute, an auto-sampler equipped DSC can run over 275 samples in a 24 hour period. This speed of analysis allows very close monitoring of processes.

However, the real advantages of fast scanning rate DSC are found in the increase in apparent sensitivity and the suppression of kinetic events due to the increased heating rate. In excipients like lactose [1], in pure or high concentration protein formulations [2], or in crystalline pharmaceuticals, [3] the ability to detect low levels of amorphous materials and to characterize its glass transition is greatly enhanced by scanning at faster rates. One use of this is in screening for solubility studies [4]. Figure 1 shows a material run to detect the presence of amorphous components which could adversely affect dissolution of the material. Scanning at 500 degrees a minute shows a distinct glass transition, which indicates amorphous material in the sample. Similar work on lactose has shown it possible to detect the presence of amorphous lactose down to fractions of a percent.

The difference in scanning rates on the visibility of transitions is shown above. At 100 C/min the Tg blends into the leading edge of the melt. At 500 C/min, it becomes a clearly detectable Tg.

Other applications of fast scanning rate DSC exploit the time dependence of kinetic events: scanning at high rates will allow one to determine the melting point and the heat capacity of materials that normally decompose on melting. Similarly, kinetic transformations like polymorphic conversions can be “trapped” so that the original concentration of the components can be measured. This allows one to be sure that the final material processed is what it should be without needing to wait for results from slower methods.

Evolved Gas Analysis: Knowing Exactly What Comes Off.

Thermogravimetric Analysis (TGA) is commonly performed to check on samples to confirm the presence of residual solvents [1]. However, TGA alone has a major disadvantage; the technique only tells you how much you have, not what it is. This lack of information is addressed by coupling the TGA to another instrument, which can identify specific compounds. Most commonly used are FTIR and MS, although recently the addition of a gas chromatograph between the TGA and the MS has been found to allow very precise identification of the volatile. All of these techniques are referred to collectively as Evolved Gas Analysis, or EVA. Individually, the abbreviations of TG-IR, TG-MS and TG-GC/MS are used. Often the TGA is replaced with a combined Differential Thermal (DT) analyzer and Thermogravimetric (TGA) analyzer. This DT/TGA, or STA (simultaneous analyzer), allows you to get thermal transitions along with the weight loss curve. Figure 3 shows data from a STA coupled to an FTIR. TG-IR with either TGA or STA is increasing commonly used for these applications. We can see the DT/TGA data gives us both the delta temperature curve and the material’s weight loss, while the IR information allows us to assign specific spectra to the various thermal events. In the case of a pharmaceutical, we would be able to look at a weight loss and confirm if it is water, acetone or benzene being evolved [5].
TG_IR of the complex decomposition of a pharmaceutical studied by TG-IR. The left hand figure show the weight loss and derivative of the drug on heating in a TGA. The right hand figure shows the the map of the offgas as analyzed in the FTIR, plotting time, wavenumber, and absorbance. Data taken from work by Dr. M. Garavaglia of PerkinElmer Italy [6]The advantage of MS and GCMS over FTIR for these applications is much lower detection limits. This can be a great advantage in situations where the weight loss consists of multiple solvents as shown in Figure 3 and the individual components are below the limit of detection by IR. Because of this sensitivity we can also look at leachates and contaminates in the sample as well as degradation products from improper storage.
An overlay of the TGA curve with the mass ions for the fragments for ethanol/methanol, acetone and water shows how TGA-MS allows the determination of the components. Data is taken from work described in the PerkinElmer Application Note on TG-MS by T. Mann et al.

Spectral DSC: Getting the Best of Both Worlds.

As discussed above, DSC is a powerful technique for studying transitions in materials from polymers to foods to pharmaceuticals, but the information is often incomplete in that structural or chemical information is inferred from the thermogram. For example, DSC allows users to see the transitions between crystalline forms in a drug or a food, but the forms must be determined by another method. Similarly, DSC can provide the energy and temperature of a change that occurs by either heating or photo-initiation, but other information, such as the specific chemistry of a particular part of the molecule must be inferred. Contra wise in Raman spectroscopy, it is easy to track a band associated with a specific bond, such as the sulfur bond in an ring system, but changes in the amorphous material from rigid to mobile, the glass transition, can be difficult to see. A simultaneous technique like DSC-NIR (near-infrared) or DSC-Raman solves these problems. Raman is often preferred for this coupling as it contains greater structural information [7] (Figure 4).
Raman and NIR spectra for acetaminophen are compared showing the greater detail in the Raman Spectra. Data collected by Dr. R. Spragg and J. Sellors of PerkinElmer ASLSCombining the two techniques allows us to apply the precise temperature control of the DSC with the ability of Raman to detect the different polymorphic structures, and obtain precise characterization of the material. As an example, consider the common painkiller, acetaminophen. Raman spectroscopy has identified the three solid-state forms of this material found under these experimental conditions (note: there is also a fourth form not seen here). Combining this technique with DSC allows us to measure precisely the temperatures at which they occur. The thermogram shows two exothermic events believed to correspond to polymorphic changes in the material as well as the endothermic melt. When DSC-Raman is performed, the Raman spectra show the conversion to form II at ~90 ºC and then to form III at ~151 ºC before melting. This greatly simplifies our understanding of the thermal behavior of the material and is much more conclusive than attempting to draw this information out of either conventional and/or modulated temperature DSC or Raman spectroscopy alone.Also of interest in combined DSC and Raman spectroscopy is the study of hydrates [8] and pseudo-polymorphs. These materials show changes caused by the loss of water or some other solvate from the molecule that appears as an endothermic transition. While Raman spectroscopy with a hot stage might determine the changes in the material accurately, without understanding the thermal data and the energy of transition, it is possible for a water loss to be seen as a polymorphic transition. In DSC, this can be detected as an endothermic event and hence become suspect.

Conclusions

Recent advances in analytical techniques allow pharmaceutical manufacturers methods to have greater control and understanding of their processes. More information can now be obtained in less time and often at lower cost.

About the Author

Kevin P. Menard, PhD, MBA, is Business Manager in Thermal Analysis at PerkinElmer.

References

1. P. Gabbott, P. Clarke, T. Mann, P. Royall, and S. Shergill, American Laboratory, 2003, August, 26.
2. D. Katayama, J.. Carpenter, M. Manning, T. Randolph, P. Setlow, and K. Menard, J. Pharma. Sci., 97, 1011, 2008.
3. M.Hurtta and I.Pitkanen, Thermochimica Acta 2004, 419(1-2), 19-29.  M. Saunders, K. Podluii, S. Shergill, G. Buckton, and P. Royall, International Journal of Pharmacetics, 2004, 274, 35
4. D. Gramaglia, B. Conway, V. Kett, R. Malcolm, and H. Batchelor, International Journal of Pharmaceutics, 2005, 301, 1-5.
5. F. Paulik, Special Trends of Thermal Analysis, J. Wiley & Sons, West Sussex, UK, 1995. T. Provder, M. Urban, and H. Barth, Hyphenated Techniques in Polymer Characterization, ACS Symposium Series, Washington D.C., 1994
6. M. Garavaglia, TG-IR Analysis of the Decomposition Products of a Drug, PerkinElmer Thermal Analysis Application Note, 74, 2005.
7. Sprunt, J.; Jayasooriya, U. Appl. Spectrosc. 1997, 51, 1410.  Harju, M.; Valkenon, J.; Jaysooriya, U. Spectrochim. Acta 1997, 47A, 1395.  Redman-Furey, N.; Bigalow-Kern, A.; Collins, W.; Cambron, R. Proceedings of the NATAS Conference, Albuquerque, NM, 2003; Vol 31, p 292.  Alexander, R.; Menard, K.; Spragg, R.; Ye, P. Proceedings of the NATAS Conference, Atlanta, GA, 2008; Vol 36, p 131. Kauffman, J.; Batykefer, L.; Tuschel, D.; Bangalore, A. J. Pharm. Biomed. Anal., in press.
8. A. Bigalow-Kern, W. Collins, R. Cambron, and N. Redman-Furey, Journal of ASTM, July August 2005, 2 (7), 42.